Load Cell & Transducer Terminology

Crosby Straightpoint

Note: Our UK offices are closed from 24th December to 1st January for the festive holiday.

  This email address is being protected from spambots. You need JavaScript enabled to view it.

 UK: +44(0)2392 484491    USA: +1 918 834 4611

Rated Capacity or Working Load Limit (WLL)
The maximum load for which the load cell is designed.

Span
Full scale output of the loadcell, usually in mV/V (milliVolts output per Volt excitation or input)

Full Scale Output (FSO)
Load cell output at rated capacity or WLL (full scale),
Usually expressed in mV/V

Zero Output
Load cell output with no load applied.
Expressed as % FSO

Best Straight Line
The straight line between Zero Output and Full Scale Output, used to determine load cell output errors and non-linearity

Non-Linearity
The deviation of load cell output from the Best Straight Line, recorded with increasing loads.
Expressed as % FSO

Hysteresis
The difference in load cell outputs from equivalent increasing and decreasing loads.
Expressed as % FSO

Maximum Error / Combined Error
The maximum error of load cell output from the Best Straight Line, considering both Non-Linearity and Hysteresis.
Expressed as % FSO

Output Symmetry
The difference in FSO from tension operation to compression operation.
Expressed as % Average FSO

Temperature Effect on Zero
The change in zero offset due to a specified change in temperature.
Expressed as % FSO / °C

Temperature Effect on Span
The change in output due to a specified change in temperature.
Expressed as % Output / °C

Creep
The change in FSO occurring over time with all conditions remaining constant.
All materials exhibit creep effects to varying degrees. Load cell creep can be minimised by matching the creep compensation of the strain gauge to the load cell element material used.
Expressed as % FSO / specified time period (usually 20 minutes).

Creep Return
Zero offset after unloading the load cell from a specified time period. Recorded after equivalent time period to which the load was applied.
Expressed as % FSO / specified time period (usually 20 minutes).

Non-Repeatability
The difference in recorded output from successive loadings under constant conditions.
Expressed as % FSO

Excitation Voltage
The safe Voltage used to obtain output values within the quoted specification.
Expressed in Vdc or Vac.

Input Resistance
The electrical resistance of the load cell’s strain gauge circuit, measured across excitation connections with signal connections open circuit.
Expressed in Ohms

Output Resistance
The electrical resistance of the load cell’s strain gauge circuit, measured across signal connections with excitation connections open circuit.
Expressed in Ohms

Insulation Resistance
The electrical resistance measured between the load cell’s strain gauge circuit and the load cell element.
Expressed in Ohms

Primary Axis
The axis of load for which the load cell is intended.

Please let us know your name.

Invalid Input

Invalid Input

Invalid Input

Please let us know your email address.

Your email does no match please enter again

Invalid Input

Invalid Input

Once you have sent this form it will instantly reach a Crosby SP representative directly, who will respond to your enquiry by email ASAP. Or if you would prefer a call back please let us know in the message.
Note: Gmail email addresses will not be accepted on this form.